Referencia de instrucciones SCPI
588 FTB-5240S/S-P/BP
Product-Specific Commands—Description
Example(s) SENS:WAV:STOP 1525.0 NM
SENS:WAV:STOP? Returns 1.525000E-006
Notes Minimum instrument sweep range is 5.0 nm.
Upon changing STOP wavelength, if necessary,
STARt wavelength will be automatically ajusted
in accordance with minimum sweep range.
See Also :SENSe[1..n][:WAVelength]:OFFSet
:SENSe[1..n][:WAVelength]:STARt
:SENSe[1..n][:WAVelength]:STOP?
:SENSe[1..n]:FREQuency:STOP
:SENSe[1..n][:WAVelength]:STOP