Referencia de instrucciones SCPI
OSA 527
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]:AUTO
Description This command controls activation of WDM
analysis OSNR calculation using auto resolution
bandwidth for all channels.
At *RST, auto resolution bandwidth is set to off
(disabled).
Syntax :CALCulate[1..n][:WDM]:OSNR:BANDwidth|BWI
Dth[:RESolution]:AUTO<wsp><Auto>
Parameter(s) Auto:
The program data syntax for <Auto> is defined
as a <Boolean Program Data> element. The
<Auto> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.