Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manuale Utente Pagina 439

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 659
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 438
Referencia de instrucciones SCPI
OSA 433
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:
WIDTh:FREQuency
Description This command sets the frequency width of the
selected WDM analysis channel.
At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
this value is set to 50.0 GHz.
Syntax :CALCulate[1..n][:WDM]:CHANnel:WIDTh:FREQ
uency<wsp><Width[<wsp>HZ]>|MAXimum
|MINimum|DEFault
Parameter(s) Width:
The program data syntax for <Width> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
element is HZ. The <Width> special forms
MINimum, MAXimum and DEFault are accepted
on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Vedere la pagina 438
1 2 ... 434 435 436 437 438 439 440 441 442 443 444 ... 658 659

Commenti su questo manuale

Nessun commento