Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manuale Utente Pagina 546

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 659
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 545
Referencia de instrucciones SCPI
540 FTB-5240S/S-P/BP
Product-Specific Commands—Description
Example(s) STAT? Must return READY
CAL:ZERO:AUTO ONCE
STAT:OPER:BIT9:COND? Keep resending this
query as long as the operation is not complete
(returned value is not 0).
CAL:ZERO:AUTO? Returns 1 (auto zero still
enabled)
Notes Zero calibration operation takes up to 5 seconds
to complete.
Auto zero calibration cannot be disabled: OFF (0)
value is valid for queries only.
See Also :CALibration:ZERO:AUTO?
:STATus?
:STATus:OPERation:BIT<8|9>:CONDition?
:CALibration[1..n]:ZERO[:AUTO]
Vedere la pagina 545
1 2 ... 541 542 543 544 545 546 547 548 549 550 551 ... 658 659

Commenti su questo manuale

Nessun commento