Referencia de instrucciones SCPI
584 FTB-5240S/S-P/BP
Product-Specific Commands—Description
Example(s) SENS:WAV:STAR 1460.0 NM
SENS:WAV:STAR? Returns 1.46000E-006
Notes Minimum instrument sweep range is 5.0 nm.
Upon changing STARt wavelength, if necessary,
STOP wavelength will be automatically ajusted in
accordance with minimum sweep range.
See Also :SENSe[1..n][:WAVelength]:OFFSet
:SENSe[1..n][:WAVelength]:STOP
:SENSe[1..n][:WAVelength]:STARt?
:SENSe[1..n]:FREQuency:STARt
:SENSe[1..n][:WAVelength]:STARt