Exfo IQS-12004B DWDM Passive Component Test System for Manuale Utente Pagina 39

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 38
Getting Started with Your DWDM Passive Component Test System
DWDM Passive Component Test System 27
Installing Hardware Components
To work in high-power mode, you will have to add a switch and an EDFA
to your standard IQS-12004B DWDM Passive Component Test System.
The minimum hardware components required for multipath testing in high
mode are:
³ Tunable laser source (IQS-2600B or IQS-2600CT).
³ Wavelength reference module (IQS-9401).
³ Power meter (IQS-1643T), one or more.
³ The KPS-AMP-C-20 fiber amplifier (EDFA).
³ A high-power GPIB-controlled switch.
IMPORTANT
In order to be able to control the EDFA, the GPIB address of the
EDFA must be smaller than the GPIB address used for the switch.
CAUTION
Never insert or remove a module while the controller or expansion
unit is turned on. This will result in immediate and irreparable
damage to both the module and unit.
Vedere la pagina 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 436 437

Commenti su questo manuale

Nessun commento