CALibration:ORLRtj
Description This command performs an ORL calibration with a reference
test jumper.
Syntax CAL:ORLR
Parameter None
Response None
Example CAL:ORLR
CALibration:ORLRtj?
Description The response to this query indicates if an ORL calibration with a
reference test jumper has been performed.
Syntax CAL:ORLR?
Parameter None
Response A boolean response representing the state of the ORL
calibration performed with the RTJ.
“1” - The ORL calibration with the RTJ is performed.
“0” - The ORL calibration with the RTJ is not performed
Example CAL:ORLR?