Exfo MaxTester Series Manuale Utente

Navigare online o scaricare Manuale Utente per Strumenti di misura Exfo MaxTester Series. EXFO MaxTester Series User Manual Manuale Utente

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 6
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 0
P/N: 1065015 July, 2013 1/6
NOTICE
This information applies only to the
power meter and VFL designed for the MaxTester
Series unit.
Installing or Removing the Power Meter and
VFL
If your unit was not equipped with a power meter (or a power meter and a
VFL) at time of your initial purchase, you can purchase one afterwards and
install it yourself.
You may also need to remove the power meter to send it to an authorized
service center for recalibration.
Note: If you prefer not to remove the power meter from your unit when a
calibration is needed, you can also send the complete unit to the service
center.
Note: The illustrations presented hereafter show a power meter and a VFL. The
appearance of your test instrument may vary slightly depending on the
model that you have purchased.
WARNING
To avoid serious injuries as well as irreparable damages both to
your unit and your power meter, ALWAYS TURN OFF YOUR UNIT
(shutdown mode, not suspend) and DISCONNECT IT FROM THE
EXTERNAL POWER SOURCE.
Always use caution during the installation or removal process.
Even if the unit is disconnected from its external power source,
the battery continues to provide power to certain areas of the
unit.
Be careful not to drop metal objects such as screws inside the
unit. These could cause a short-circuit leading to a fire or an
explosion.
Vedere la pagina 0
1 2 3 4 5 6

Sommario

Pagina 1 - Series unit

P/N: 1065015 July, 2013 1/6NOTICEThis information applies only to thepower meter and VFL designed for the MaxTester Series unit.Installing or Removing

Pagina 2 - P/N: 1065015 July, 2013 2/6

P/N: 1065015 July, 2013 2/6CAUTIONElectrostatic discharge (ESD) damage can cause complete or intermittent equipment failures.  Always use an ESD-prev

Pagina 3

P/N: 1065015 July, 2013 3/6To install a power meter:1. Press and hold the On/Off button a few seconds until the unit beeps once to put it in shutdown

Pagina 4 - Writings

P/N: 1065015 July, 2013 4/68. By holding the power meter by its faceplate, position it so that the writings are right-side up.9. Carefully align the p

Pagina 5

P/N: 1065015 July, 2013 5/6To remove the power meter for recalibration:1. Press and hold the On/Off button a few seconds until the unit beeps once to

Pagina 6 - P/N: 1065015 July, 2013 6/6

P/N: 1065015 July, 2013 6/67. By holding the power meter by its faceplate, gently pull it out of the bay completely.8. Place the power meter in an ant

Commenti su questo manuale

Nessun commento