Exfo MaxTester 635 Manuale Utente Pagina 169

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 168
Noise Tests
Copper and VDSL2/ADSL2+ Multi-play Test Set 161
NEXT
NEXT
The NEXT test measures the Near End Crosstalk (NEXT) between 2 pairs of
cables. The test transmits a signal on the T/A - R/B pair, and measures the
received crosstalk on T1/A1 - R1/B1 pair.
Spot
Spot is a continuous mode of operation measuring
the crosstalk for a specific frequency selected by you,
similar to the method used by the Send Tone test. The
Spot function displays and updates results in real
time, until you stop the test. The last valid results then
remain displayed.
Current Group is the current test group name, for
example Noise Tests.
Current Test displays the test being measured.
Here you can switch to another Noise test from
the given list. However, this action will terminate
the ongoing measurement test.
Frequency allows you to set the frequency value for the Near End
Crosstalk (NEXT).
Frequency Step lists a choice of frequency step values.
Termination Resistance lists a choice of resistance values for the
termination of the signal.
Show hookup diagram on test setup - If Yes, the Confirmation page is
prompted before every NEXT test is started. Select No to hide the
diagram next time you run the test.
Vedere la pagina 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 354 355

Commenti su questo manuale

Nessun commento