SCPI Command Reference
Tunable Laser Source 75
Product-Specific Commands—Description
:INITiate[1..n][:IMMediate]
Description This function starts the sweep program, which
performs a sweep from a start wavelength to a
stop wavelength, as specified in the sweep
parameters.
At *RST, a sweep in progress is aborted.
Syntax :INITiate[1..n][:IMMediate]
Parameter(s) None
Example(s) SOUR:WAV:SWE:REP NCON
SOUR:WAV:SWE:DIR BOTH
SOUR:WAV:SWE:CYCL 12
SOUR:WAV:SWE:STAR 1523.45NM
SOUR:WAV:SWE:STOP 1600.45NM
SOUR:WAV:SWE:SPE 25.56NM
SOUR:WAV:SWE:MODE CONT
TRIG SEQ:INP DISABLED
INIT:IMM
Notes The sweep can be started even if the optical laser
is off.
See Also ABORt[1..n]
SOURce[1..n]:WAVelength:SWEep:STATe?