SCPI Command Reference
Tunable Laser Source 73
Product-Specific Commands—Description
:CALibration[1..n]:WAVelength
Description This command introduces an offset to correct a
wavelength deviation as measured by an
external wavelength meter.
INITiate[:IMMediate] is an event and cannot be
queried as there is no state associated with it.
Syntax :CALibration[1..n]:WAVelength<wsp><Readout
Value>
Parameter(s) ReadoutValue:
The program data syntax for <ReadoutValue> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
This parameter is the wavelength value in meters
read by a calibrated wavelength meter.