Exfo IQS-2600B Tunable Laser Source for IQS Platforms Manuale Utente Pagina 107

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 106
SCPI Command Reference
Tunable Laser Source 101
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:
STARt
Description This command allows you to specify a start
wavelength for the sweep.
At *RST, the start wavelength is 1520 nm.
Syntax :SOURce[1..n]:WAVelength:SWEep:STARt<wsp
><StartWavelength[<wsp>M|HZ]>|MAXimum
|MINimum|DEFault
Parameter(s) StartWavelength:
The program data syntax for <StartWavelength>
is defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> elements are: M|HZ. The
<StartWavelength> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.