Exfo FVA-3150 Variable Attenuator Manuale Utente Pagina 65

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 64
Maintenance
Variable Attenuator 59
Recalibrating the Unit
Recalibrating the Unit
Manufacturing and service center calibrations are based on the
ISO/IEC 17025 Standard, which states that calibration documents must not
contain a recommended calibration interval, unless this has been
previously agreed upon with the customer.
Validity of specifications depends on operating conditions. For example,
the calibration validity period can be longer or shorter depending on the
intensity of use, environmental conditions and unit maintenance. You
should determine the adequate calibration interval for your unit according
to your accuracy requirements.
Under normal use, EXFO recommends calibrating your unit every year.
Vedere la pagina 64
1 2 ... 60 61 62 63 64 65 66 67 68 69 70 ... 114 115

Commenti su questo manuale

Nessun commento