SCPI-Befehlsreferenz
OTDR 377
Product-Specific Commands—Description
:CONFigure[1..n]:ACQuisition:
DURation
Description This command specifies the duration that will be
used for the next acquisition.
*RST reverts this setting to default value.
Syntax :CONFigure[1..n]:ACQuisition:DURation<wsp>
<Duration>|MAXimum|MINimum|DEFault
Parameter(s) Duration:
The program data syntax for <Duration> is
defined as a <numeric_value> element. The
<Duration> special forms MINimum, MAXimum
and DEFault are accepted on input.
MINimum allows to set the instrument to the
lowest supported value.
MAXimum allows to set the instrument to the
highest supported value.
DEFault allows the instrument to select a value
for the <Duration> parameter.
Sets the acquisition duration, in seconds.
Example(s) CONF:ACQ:DUR? Ex.: Returns 15
CONF:ACQ:DUR 10
CONF:ACQ:DUR? Returns 10
See Also FETCh[1..n]:DURation?
FETCh[1..n]:ASETting:DURation?