Exfo FTB-7000 OTDR Series for FTB-500 Manuale Utente Pagina 320

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 319
Description of Event Types
306 OTDR
Pass/Fail Tests
Not All Event Types Are Tested
In this situation, the only thing that we can clearly know is when the loss
has a Pass status. If the global event loss is less than or equal to the
smallest threshold value (a value that is tested, of course), we are sure that
the merged event status is Pass. Otherwise, we cannot know, so the status
of the event is Unknown.
In our example, if we suppose that the you chose not to test non-reflective
losses, then the analysis would be done as shown below:
Vedere la pagina 319
1 2 ... 315 316 317 318 319 320 321 322 323 324 325 ... 481 482

Commenti su questo manuale

Nessun commento