Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manuale Utente Pagina 571

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 570
SCPI Command Reference
OSA 565
Product-Specific Commands—Description
:SENSe[1..n][:WAVelength]:STARt?
Description This query returns a value indicating either the
current or the minimum/maximum instrument
sweep start wavelength.
At *RST, this value is set to 1525.0 nm.
Syntax :SENSe[1..n][:WAVelength]:STARt?[<wsp>MAXi
mum|MINimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Start>
Vedere la pagina 570
1 2 ... 566 567 568 569 570 571 572 573 574 575 576 ... 637 638

Commenti su questo manuale

Nessun commento