SCPI Command Reference
Optical Spectrum Analyzer 505
Product-Specific Commands—Description
:SENSe[1..n]:WAVelength[:RANGe]:
STOP
Description This command sets the stop wavelength for
acquisition range.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is set to 1565.0 nm (C-band
upper limit).
Syntax :SENSe[1..n]:WAVelength[:RANGe]:STOP<wsp>
<Stop[<wsp>NM]>|MAXimum|MINimum
Parameter(s) Stop:
The program data syntax for <Stop> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
element is NM. The <Stop> special forms
MINimum and MAXimum are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
Maximum wavelength for acquisition range.
Range is device-dependent.