Справка по командам SCPI
OSA 457
Product-Specific Commands—Description
Response(s) Center:
The response data syntax for <Center> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.
The <Center> response corresponds to either the
current or the MINimum/MAXimum channel
center wavelength in meter.
Example(s) CALC:CHAN:DEF "CWDM_7",1450.0 NM
CALC:CHAN:SEL "CWDM_7"
CALC:CHAN:CENT:WAV? Returns
1.45000E-006
See Also :CALCulate[1..n][:WDM]:CHANnel[:DEFine]
:CALCulate[1..n][:WDM]:CHANnel:CENTer[:WAV
elength]
:CALCulate[1..n][:WDM]:CHANnel:CENTer:FRE
Quency?
:CALCulate[1..n][:WDM]:CHANnel:SELect
:CALCulate[1..n][:WDM]:CHANnel:
CENTer[:WAVelength]?