Exfo FLS-5800 CD/PMD Analyzer Source Manuale Utente Pagina 84

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 106
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 83
Instrucciones de control remoto
78 FLS-5834A
Instrucciones IEEE 488.2: Descripción
*OPC?
Description The *OPC? query allows synchronization
between the instrument and an external
controller by reading the Output Queue or by
waiting for a service request on the Message
Available (MAV) bit in the Status Byte Register.
The *OPC? query causes the instrument to place
an ASCII character, 1, into its Output Queue
when the device completes all pending
operations. A consequence of this action is that
the MAV bit in the Status Byte Register is set to
state 1.
Syntax *OPC?
Parameter(s) None
Response Syntax <Acknowledge>
Response(s) Acknowledge:
The response data syntax for <Acknowledge> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Acknowledge> response is a single
ASCII-encoded byte corresponding to 1.
The receipt of an <Acknowledge> response
indicates that all pending selected device
operations have been completed.
Example(s) *OPC? Return 1
See Also *OPC
*WAI
Vedere la pagina 83
1 2 ... 79 80 81 82 83 84 85 86 87 88 89 ... 105 106

Commenti su questo manuale

Nessun commento